Non-Magnetic Drill Collars Functions & Selection

Non-magnetic drill collar is made from specially treated steel alloys so that it does not affect magnetic surveying instruments. Their purpose is to isolate directional survey instruments from magnetic distortion due to the steel drill string.

Non Magnetic Drill Collar Function

The magnetic survey instrument is influenced by the presence of a magnetic field around the tool: ferromagnetic material in the formation, drill pipe, drill collar, or other bottom hole assemblies. This influence can’t be completely eliminated, but can be reduced to minimize interference with the earth’s magnetic field. The survey instrument is placed inside a non-magnetic drill collar so that the survey instrument can measure the earth’s magnetic field. More than one non-magnetic drill collar may be used to minimize magnetic interference from the drill string and bottom hole assembly.

Material

Non-magnetic drill collars are manufactured from a material called Monel. This Monel contains a group of nickel alloys, mainly formed of nickel (ranged 52 to 67%) and copper, with slight amounts of silicon, iron, carbon, and manganese. It is much stronger than a pure nickel. In addition to being non-magnetic, Monel alloys have a high resistivity to corrosion by high aggressive agents, including rapidly flowing seawater. They can be fabricated readily by hot- and cold-working, machining, and welding.

Non-Magnetic Drill Collar Specification & Selection

Non-magnetic drill collars are also called Monel drill collars. The location of the survey tool inside the collar and the number & length of the non-magnetic drill collars should be carefully selected. Their selection is a function of geographical location, inclination, azimuth, and bottom hole assembly.

Earth’s Magnetic Field

The number and length of the non-magnetic drill collars needed depend upon the geographical location of the well. The location will dictate the horizontal intensity of the earth’s magnetic field and hence the quality of the survey reading. The horizontal intensity is relatively small close to the earth’s poles and is higher near the equator. A location where the horizontal intensity is low would require a greater length of non-magnetic drill collars as compared to where the intensity is high.

The horizontal intensity is a function of magnetic dip angle and is given by

Horizontal Intensity = Magnetic Field Strength X cosine (Magnetic Dip Angle),

The magnetic dip angle is the angle that a magnetic needle makes with a horizontal plane at any location on the earth. This angle is close to 0o (actually -20o to 20o depending upon location) near the equator and 90o at each of the magnetic poles, and so, the horizontal intensity is highest near the equator and smallest at the magnetic poles. As an example, the magnetic field strength and the dip angle are 57,500 gammas and 80.6 degree on the North slope of Alaska. The same quantities are 48,500 gammas and 59.7 degrees in the Gulf Coast in US. Thus, the horizontal intensity of the earth’s magnetic field is only 9,391 at Alaska, but 24,470 gammas in the Gulf Coast. Hence, it will take fewer non-magnetic drill collars to survey wellbore direction in the Gulf Coast as compared to the survey on the north slope of Alaska. The stronger the horizontal intensity of the earth’s magnetic field, the less the interference from the drill string.

World Zonal Map

A number of years ago, Shell created a zonal map of the world to describe the horizontal intensity of the earth’s magnetic field. The intensity is strongest in Zone I, and progressively weaker in Zone II and Zone III. This zonal map is shown in Figure 1. Wellbore surveying in Zone I can be performed with fewer non-magnetic drill collars, whereas surveying in Zone III will require more non-magnetic drill collars to accurately determine the direction of a well.

Magnetic Zonal map For Drill collars

Figure 1. Map Showing the Geographical Location of the Zones for Use in Selecting Non-Magnetic Drill Collar Lengths

Inclination & Azimuth

The number of non-magnetic drill collars also depends on the inclination and direction of the well. The higher the inclination angle of the wellbore, the larger is the number of drill collars required. As the direction of the wellbore is changed, additional non-magnetic drill collars may be required. The surveying of a wellbore pointing to the magnetic North or South would require less number of drill collars, whereas a wellbore oriented towards the East or West will require additional collars.

The length of the non-magnetic drill collars for the three zones can be approximated using Figure 2 through Figure 4. Figure 1 is first used to determine the zone where the well is located. Then, the chart corresponding to this zone is utilized to select the length of non-magnetic drill collars and the compass spacing within the collars. Note that as the inclination increases for a given hole direction, the length of the non-magnetic collars required increases. Also, the collar length required increases as the direction progresses toward east or west.

Data Chart for Zone I
Figure 2. Data Chart for Zone I
  • Use 18′ collar in area below Curve A
  • Use 25′ foot collar in area below Curve B
  • Use 30 foot collar in area below Curve C
  • Use tandem (18′ + 25′) collars in areas above Curve C
Compass Spacing:18′ collar:1′ to 2′ below center
25′ collar:2′ to 3′ below center
30′ collar:3′ to 4′ below center
Tandem (18′ + 25′):center of bottom collar
Data Chart for Zone II
Figure 3. Data Chart for Zone II
  • Use 30′ collars in the area below Curve
  • Use 60′ collars in the area below Curve B with packed bottom hole assembly
  • Use 60′ collars in the area below Curve C with near bit stabilizer only
  • Use 90′ collars in the area above Curve C
Compass Spacing:30′ collar:3′ to 4′ below center
60′ collars:At center (Curve B)
60′ collars:8′ to 10′ below center (Curve C)
90′ collars:At center
Data Chart for Zone III
Figure 4 Data Chart for Zone III
  • Use 60′ collars in the area below Curve A with packed bottom hole assembly
  • Use 60′ collars in the area below Curve B with near bit drilling stabilizer only
  • Use 90′ collars in the area below Curve C with any bottom hole assembly
Compass Spacing:60′ collarsAt the center (Curve A)
60′ collars:8′ to 10′ below center (Curve B)
90′ collars:At center

Total Magnetic Field Effect

The non-magnetic drill collar length and compass spacing are not determined by the earth’s magnetic field alone. It is determined by the total magnetic field which is the vectorial sum of the earth’s magnetic field and the magnetic field caused by the magnetized portions of the drill string. Since the drill string is much longer than its diameter, it can be analyzed as though it were a point source. In most cases, there is a North pole (+) above the non-magnetic drill collar and a South pole (-) below the non-magnetic drill collar. The pole strengths determine the non-magnetic collar length. The stronger the pole, the greater the length required. The actual pole strength depends upon the specific bottom hole assembly component. Pole strengths can vary significantly even when comparing the same size collars. The selection charts have assumed constant pole strengths; therefore, care should be exercised in using the charts.

Compass Spacing For Non-Magnetic Drill Collars

Generally, the upper pole strength (North) is greater than the lower pole strength (South) mainly due to the length of the drill string. The effect of the magnetic forces caused by the drill string cannot be completely eliminated; however, it can be minimized. The point at which the magnetic field has a minimum effect is the optimum placement for the compass. With short bottom hole assemblies (weaker magnetic pole), the optimum compass spacing is below the center of the non-magnetic collar length. With longer bottom hole assemblies (such as a hold assembly or downhole motor), the optimum spacing is closer to the center of the non-magnetic collar length. Non-magnetic drill collars are manufactured from a solid piece of non-magnetic steel and the ID is bored from both ends. The bore may not match exactly in the center and can cause a slight magnetic field. It is usually best not to put the magnetometers in the exact center of the non-magnetic drill collar.

Conclusion

Nonmagnetic drill collars are made from Monel and the Earth’s magnetic field is undisturbed by their presence. An accurate reading of the direction of the well can therefore be obtained. The number of collars that are required depends on the magnetic latitude and hole direction. The compass is actually measuring the horizontal component of the Earth’s magnetic field. Where the magnetic field lines are steeply dipping and
the hole direction is close to the East-West axis the horizontal component is small, and so more non-magnetic collars must be used (Figure 5)

Non-magnetic drill collars selection chart
Figure 5

Leave a Comment